Evaluation of the Accuracy of Different Apex Locaters with 45° Apical Root Resorption

著者

  • Ahmed I Al-Jobory
  • Mohammed Ayad Taha College of Dentistry, Tikrit University.
  • Mohammed Jabbar Atyiah

##semicolon##

https://doi.org/10.5195/d3000.2024.765

##semicolon##

Apex Locater Devices##common.commaListSeparator## Working Length##common.commaListSeparator## Electronic Working Length Test##common.commaListSeparator## 45° Root Resorption

要旨

Aim: This study was aimed to calculate the working length (WL) of permanent teeth with simulated 45° resorption of the apical part of the root using four electric apex locaters (EALs): NSK, Woodpecker III, Woodpecker V, and Eighteeth.

Methods: Twenty maxillary anterior single-rooted teeth were removed. Following tooth preparation to offer access to the root canal and replicate the 45° apical root resorption, each tooth underwent a microscope-assisted working length determination process. The four apex locaters measured each individual tooth and calculated electronic working lengths.

Results: No discernible change existed between all four apex locater devices.

Conclusion: NSK, Woodpecker III, Woodpecker V, and Eighteeth provide virtually the same measurement for 45 apical root resorption in single rooted teeth.

##submission.downloads##

出版済

2024-11-22

巻号

セクション

Adults & the Elderly